Milwaukee SE600 - Combo pH / EC / TDS Probe
USE ORIGINAL MILWAUKEE REPLACEMENT GRADE.
The Milwaukee allows you to easily and accurately test pH, conductivity and total dissolved solids (TDS) using an integrated probe.
The Milwaukee SE600 pH/EC/TDS combination electrode features a female DIN plug and 1 meter cable.
DESIGN FEATURES
Dual Connection: The dual connection design reduces the risk of clogging.
Gel-filled: The use of a gel eliminates the need to maintain electrolyte levels.
Resin Body: The resin body is highly durable, chemical resistant and protects against accidental damage.
USE AND MAINTENANCE
The SE600 probe contains both a pH and an EC electrode. Both require maintenance.
pH
All pH sensors have a limited lifespan and the response time gets longer as they age. Properly maintaining your pH pin tester will help you get fast, accurate results for longer. It is necessary to clean and condition the pH electrode before use!
TOP 4 TIPS FROM MILWAUKEE
- Always store with the cap full of buffer or storage liquid not water.
- Clean your tester regularly by immersion in cleaning solution.
- Calibrate after cleaning and/or storing before testing to correct for changes in the pH and EC sensor.
- Gently stir the tester into the sample for faster results.
EC
EC sensors are more robust, but can still be affected by residue buildup.
EC electrodes may start to release nutrients over time. Rinse the probe with fresh tap water after each use to prevent build-up.
CONVERTING EC TO TDS
The Milwaukee EC pen, like all conductivity meters, provides a measurement of electrically charged ions in a solution and is an absolute measurement of conductivity.
EC is often converted to TDS in parts per million (ppm). There are several scales, such as the 500 scale, the 650 scale, and the 700 scale. However, the true PPM can only be determined by chemical analysis.
Your mS/cm reading can be converted to TDS (ppm) by approximation:
2,000 µS/cm = 2.0 mS/cm = 1000 ppm on .50 factor scale
2,000 µS/cm = 2.0 mS/cm = 1400 ppm on 442 scale or .70 factor scale
Specifications
- Housing: Polypropylene
- Reference: Single, Ag/AgCl
- Connection: Cloth
- Electrolyte: Gel
- pH range: 0 - 13 pH
- Temp Range: 20 - 40 C (68 - 104 F)
- Tip Shape: Spherical
- Max. Pressure: 0.1 bar
- Length: 3.8 inches
- Cable Length: 7 pole / 3.3 foot
- Connection: DIN
- Ideal to use with: